this post was submitted on 13 Oct 2023
4 points (100.0% liked)

Ask Electronics

3316 readers
1 users here now

For questions about component-level electronic circuits, tools and equipment.

Rules

1: Be nice.

2: Be on-topic (eg: Electronic, not electrical).

3: No commercial stuff, buying, selling or valuations.

4: Be safe.


founded 1 year ago
MODERATORS
 

Hi!

I've asked Siglent support but after a couple of responses the thread went cold. Maybe I'm being dumb but I've noticed that there's a frequency (low, around 100Hz) where scope response changes a bit. Below and above it square input looks square. Right about it square input looks slanted.

I tried to do a very slow "sweep" and there's small but visible change in the envelope. So, e.g. with a constant 600mV p2p input lower frequencies measure exactly that while higher ones measure 612mV, so ~2% diff.

Terminated 50Ohm cables (not that it matters at such a low freq) to be sure. Latest firmware, after full self-cal. Siggen itself seems allright, I have an ancient Tek scope and the siggen output looks Ok there with same input/same cables. Scope seems happy and fully functional otherwise, few years old though, out of warranty.

Has anyone else seen anything like that? Is this a normal behavior within the expected margin of error?

Thanks!

you are viewing a single comment's thread
view the rest of the comments
[โ€“] [email protected] 1 points 1 year ago* (last edited 1 year ago) (1 children)

Yup, seen that for sure.

Did you try turning the little adjustment knob in your probe to calibrate it? Sometimes needs a small screwdriver. Here's a reference:

https://www.elecrow.com/download/HowToCalibrate10xProbe.pdf

What I'm referring to is labelled 'Cap Trimmer'. The document also has some waveform images that match your problem.

I have a Siglent and it looked like this at the dealership, then they adjusted the probe a bit, and then it was 100% fine.

[โ€“] [email protected] 1 points 1 year ago

Thanks but unfortunately this is not a probe cal problem. Note that the issue is seen even in a 50Ohm environment, no probes.